INTRODUCTION

Optical flow monitoring systems are used for contiuous gas analysis with corrosive sample gases, to
have the possibility of using flowmeters with corrosion resistant floating balls out of e.g. glass.
The optical flow monitoring systems FA1-H-.. are an alternative to the optical flow monitoring systems
FA-1 with light fork barrier. These units can be used in heated gas conditioning systems up to 180°C
also with non metallic or very small floating balls.


7

APPLICATION

The optical flow monitoring unit with light guides is used in gas analysis on heated flowmeters with
temperatures > 60 °C. The measuring cones must be transparent. Thanks to optical scanning, very
low flow quantities can be recorded even in case of flowmeters with non-metallic or very small (1 mm)
floats.


8

TECHNICAL DATA

Sensor Head
FA2-H

Light guide
FO1

Light guide
FO3

Light guide
FO2

pre-amplifier
K-FA-H

Part Number

02 E 4002

02 E 4060

02 E 4063

02 E 4065

02 E 4010

Function
monostable

1x

2x

1x

Measuring tube

5-14 mm

Light guide length

600 mm

900 mm

1200 mm

Dimensions
W x D x H

24 x 40 x 24

ø 6 mm

80 x 160 x 55

Weight

90 g

230 g

330 g

420 g

560 g

Material

anodized
aluminium

glass fibre, SS 316Ti, brass-chromed

Polycarbonate
combustibility
class
VO (UL94)

Assembly

Fixed with clamping screw

Wall mounting

Cable entry

terminal range
5mm - 10mm

Electrical
connection

terminals
2,5mm

2

Supply voltage

12V DC internal
from FA-1..

Electrical standard

EN 61010

Operating
temperature

- 25°C to + 180°C

-25°C to + 60°C

Storage
temperature

- 25°C to + 70°C

System of
protection

IP65 EN 60529