beautypg.com

Figure 2, Electrical connection of the pre-amplifier k-fa-h, Figure 3 – M&C TechGroup FA1-H Operator's manual User Manual

Page 12: Electrical connection fa1-h

background image

12

Gas sampling and gas conditioning technology

5-6.15ME

For the electrical connection, the following steps have to be carried out (see also figure 2+3):

Loosen the 4 lid screws and remove the lid of pre-amplifier K-FA-H.

Lead the connection cable through the respective clamping screw.

The electronic controller FA-1.. has to be connected on clamping block X1, terminals 5 =
yellow, 7 = white and 8 = brown.


Further electrical connection of the electronic controller FA-1.. is described in manual 5-6.10ME,
chapter 13.

Figure 2

Electrical connection of the pre-amplifier K-FA-H

Figure 3

Electrical connection FA1-H

8 7 6 5

18 17 16 15

1 2 3 14 11 12

L N P

E

br ws

ge

br ws

ge

FA-K-H

FA1.1
and
FA1.4

NC