TECHNICAL DATA

Technical Data

FM-2K

FM-200K-H

FM-200K-H/FA

Part-No.

09F2105

09F2505

09F2555 **

Heated / needle valve in the
inlet

no

yes

yes

Optical flow control, mono-stable **

no

no

fix between 20 - 50%

of scale

Standard measuring range

25-250 l/hr air, 1.2 bar, 180°C

Measuring ranges available

Scale: 1.6-16, 6-60, 10-100, 25-250, 50-500, 80-800 l/hr air,
1.2 bar, 180°C

Operating pressure / operating
temperature

max. 2 bar / max. +180°C

Ambient temperature

-40 to +180°C

-25 to +60°C

-25 to +50°C

Storage temperature

-40 to +80°C

-25 to +80°C

Gas connections

¼“ NPT i

Tube fittings Swagelok ø6mm, optional ø1/4“

Temperature controller

-----------------------------

Capillary thermostat, integrated in electrical

connection box with high temperature limiter

and low temperature alarm contact

Operating temperature

-----------------------------

adjustable from 0 to 180°C, set in factory at

180°C

Temperature alarm contact

-----------------------------

alarm point

T -30°C to T

Set

change over contact, voltage free; contact

rating 250V, 3A

, 0,25A=

Power supply

-----------------------------

230V 50Hz, 220VA optional 115V 60Hz (a)

Electrical connections

-----------------------------

Terminals 2,5 mm², 2x Cable glands PG13

Protection / Electrical
standard

-----------------------------

IP 54 EN60529 / EN 61010, EN60519-1

Dimensions

50 x 290 x 50mm (w x

h x d)

250 x 500 x 140mm (w x h x d)

Weight

0.8 kg

8 kg

8.5 kg

Dead volume / Mounting

approx. 6 cm³ / wall mounting

Material of sample contacting parts

glass, stainless steel 316, PTFE

** The optical flow alarm is only equipped with pre-amplifier K-FA-H, the electronic controller must be ordered separately



.

0

10

20

30

40

50

0

150

300

450

600

750

900

1050

Nl/hr