DESCRIPTION
The M&C liquid sensor KS2 / KS2.Ex works on the principle of capacitive measurement and is
suitable for non-conductive and conductive media. A pre-amplifier is integrated in the sensor housing
and is connected with the necessary external electronic controller via 2- resp. 3-wire. For the sensor
type KS2 the required electronic controller is available in various versions, FA1.1 or FA1.4 and is
described in a separate data sheet. The M&C liquid sensor KS2.Ex for applications in hazardous
areas and media should only be applied in connection with electronic controller WE77.Ex1, KFA6-
SR2-Ex1.W or an electronic with the same performance data (see chapter 9.4). For electronics with
the same performance data there is no guarantee for problemless operation. The M&C KS2 / KS2.Ex
liquid sensor is constructed in such a way that any droplets of liquid in the sample gas are attracted
directly to the active sensor surface. Even small liquid droplets will trigger a sure and rapid alarm. The
sensors can be mounted with the 16mm
stainless steel body in the GL-25 connector of the
Universal filter
F..-..-D or the condensate pot TG1 or in the flow chamber LS/KS.
max.
TG 1
KS 2
Figure 2
KS2.. liquid sensor for level monitoring in the condensate pot type TG1
In the event of condensate breakthrough, the filter acts as a buffer vessel preventing immediate liquid
penetration.
42
58
ø5
13
305
*
G1/4" i.
G1/4" i.
G1/4" i.
* constructional size 410mm
KS 2
Figure 3
Liquid alarm-sensor KS.2 / KS2.Ex in a Universal filter F..-..-D