Calibrating the output current, Gain calibration, Offset calibration – Xantrex Technology XTR 850 Watt User Manual
Page 194: Calibrating the output current –6, Gain calibration –6 offset calibration –6

Calibration and Troubleshooting
6-6
975-0200-01-01
The SCPI Command (s) for these instructions are:
[:]CALibration[ reading in volts>} Calibrating the Output Current Gain Calibration To perform gain calibration: 1. Disconnect the power supply from the load. 2. Connect the output terminals to a load representing a short circuit using a precision ammeter. 3. Turn both the power supply and the load On. 4. Set the nominal level to 90% of nominal current level. 5. Turn on the output power. 6. The power supply starts up in Constant Current mode. 7. Read the current value on the ammeter display. 8. Type SCPI command with current value as parameter “data”. For example, you would type :CAL:OUTP:CURR 5.57 if you had Offset Calibration Offset calibration of the power supply provides the best accuracy in low- To perform offset calibration: 1. After performing gain calibration, set the current to 10% from the nominal. 2. Read the current value on the ammeter display. 3. Type SCPI command with the current value as parameter "data". For example, you would type :CAL:OUTP:CURR 0.50 if you had Important: For best results, both calibrations may be repeated several
an XTR 150-5.6 calibrating with the current set point set to 5.6 A.
range current.
an XTR 150-5.6 calibrating with the current set point set to 0.56 A.
times.