Set temperature status negative transition filter, Scpi error/event queue, Scpi error/event queue –46 – Xantrex Technology XTR 850 Watt User Manual
Page 168

Remote Operation
5-46
975-0200-01-01
Query Format:
[:]STATus[
:PTRansition?
Set Temperature Status Negative Transition Filter
SCPI command:
[:]STATus[
:NTRansition
Query Format:
[:]STATus[
:NTRansition?
SCPI Error/Event Queue
The error/event queue contains items that include a numerical and textual
description of the error or event. Querying for the full queue item (for
example, with SYSTem:ERRor[:NEXT]?) will return a response with
the following syntax:
Description>; The The second parameter of the full response is a quoted string containing an The maximum string length of As errors and events are detected, they are placed in a queue. This queue -350,"Queue overflow
32767]. All positive numbers are instrument-dependent. All negative
numbers are reserved by the SCPI standard with certain standard error/
event codes described in the SCPI 1999 standard document. The value,
zero, is also reserved to indicate that no error or event has occurred.
is first in, first out. If the queue overflows, the last error/event in the
queue is replaced with error: