beautypg.com

Xerox Digital copier printers User Manual

Page 85

background image

6

6-7

Understanding calibration

6.

Configure the densitometer to factory defaults and save the settings. (See the
densitometer manual for details.)

Factory default settings for the X-Rite DTP32 are:

7.

Turn on the PowerBook.

8.

In the PowerBook Setup Control Panel, select Normal for Modem Controls.

9.

Calibrate the densitometer, as described below.

T

O

CALIBRATE

THE

X-R

ITE

DTP32:

1.

On the X-Rite DTP32, simultaneously press the two buttons marked MENU.

The words

MAIN MENU

appear in the display.

2.

Press the p1 key until p2 appears.

3.

Press the cal key.

CALIBRATION

appears in the display.

4.

Press the den key.

CALIBRATING TRANSMISSION

is displayed, followed by the words

INSERT CAL STRIP

.

5.

Position the strip guide at 5.

6.

Insert the arrow end of the Auto-Cal strip into the entrance of the X-Rite DTP32 until
the roller starts pulling the strip.

Page 1

LANG (ENG)

TONE (loud)

TYPE (ser)

Page 2

BAUD (9600)

HAND (xon)

AXMT (on)

Page 3

DPT (on)

SEP (spc)

DLIM (crlf )

Page 4

DEF (off )

X10 (off )

DAP (off )

Page 5

M/M (max)

LOCK (off )

Strip guide