3 maintenance and inspection, 1 preventive maintenance schedule list – Kodak 2400DSV User Manual
Page 68
![background image](/manuals/159164/68/background.png)
Disassembly and Cleaning
5
61
5.3
Maintenance and Inspection
5.3.1 Preventive Maintenance Schedule List
5.3.2 Details of readjustments needed when parts are replaced
Replacement of the following parts calls for rechecking, readjustments, or
resetting of certain items.
Follow the corresponding procedure given below whenever they have
been replaced.
Parts
Cleaning Cycle
Replacement Cycle
Refer to
Reader and Scan Path Mirrors
When image
trouble occurs
-------
Friction Plate
-------
300,000 scans
Bushing
-------
300,000 scans
Projection Lamp
-------
When “L2” displays
*Replacement cycle is the number of times of scan.
Name
Readjustments /Resetting/Rechecking
PWB-A
1. Shading Correction “b6”
2. Auto Exposure Coarse Adjustment
PWB-BB
1. Shading Correction “b6”
2. Scan Speed Adjustment
Turn off all DIP switches (DSW1) on the new board.
Make the function mode setting (S,y,o) on the new board.
(“About the Function Mode Setting Label” on page 82)
PWB-SS
Make the same SCSI ID setting for Rotary Switch (RSW1) on the new
board as those of the old board.
PU2
Projection Lamp Voltage Adjustment “F7”
Refer to “Function Mode and Adjustment” on page 77 for the adjustment procedure.