Copper wire site requirements, Site wiring requirements – Siemens ADVANCE NETWORK COMMUNICATIONS BOARD/DATANET 2000592-001 User Manual
Page 19
Site Wiring Requirements,
Continued
Table 1-1: Recommended
DataNET Cable,
Continued
Cable Type
Vendors & P/N
Installation Notes
Twisted Pair Copper
Wire used in SEA
2100 Installations
SEA P/N C00128
Belden P/N 9462
Not Recommended for
New Installations
Nominal capacitance
155 pF/m (50 pF/ft.)
Max distance from
analyzer to ANG or
between hubs 225
meters (750 ft)
Copper Wire Site
Requirements
The following information pertains when using copper wire.
See Page 11, CE Installations.
CE Installations
Use approved Conduit wiring method. Conduit may contain multiple
cables. Adhere to all Federal and local electrical code requirements.
NEC Div 2 or Non-
Hazardous Rated Areas
Cable Type: Refer to Table 1-1.
Use Cable Gland wiring method. Each cable gland allows entry of one
cable.
Cenelec Zone 1 or Zone 2
Rated Areas
Cable Type: Recommend Belden 9182 or armored Belden 9182 or
equivalent. The proper cable gland must be used to connect the armor or
shield to the enclosure. Other cables with individually shielded pairs can
be used, but decreased distance is allowed (typically ~800-1000 feet for
typical instrumentation cable).
2000592-001
Installation
• 9