Datanet channels, Site wiring requirements – Siemens ADVANCE NETWORK COMMUNICATIONS BOARD/DATANET 2000592-001 User Manual
Page 18
Site Wiring Requirements,
Continued
DataNET Channels
The DataNET channel connections can be either copper wire or fiber
optic cable. The wiring method and cable type for DataNET is dependent
upon the site requirements. For cable requirements see Table 1-1.
The following Table should be used when evaluating or selecting cable
for an existing or new DataNET installation.
Table 1-1:
Recommended DataNET
Cable
Cable Type
Vendors & P/N
Installation Notes
Fiber Optic
MM 62.5/125; 2 Fiber
Conductors
SEA P/N 1270002-002
Industry Available from
Multiple Vendors
New Installation as
an alternative to
copper cable
Max distance from
Analyzer to ANG or
between Hubs
1520 meters (5000
ft)
2- Conductor Copper
Wire
SEA P/N 1686002-001
Belden P/N 9182
For
Belden Armored Cable
order
SEA P/N 1680000-002
Recommended for
Single Channel
Max distance from
Analyzer to ANG or
between Hubs
1520 meters (5000
ft)
4-Conductor Copper
Wire
SEA P/N 1681000-003
Belden P/N SSD1743
Recommended for
Dual Channel
Max distance from
Analyzer to ANG or
between Hubs
1520 meters (5000
ft)
8
• Installation