Fluke 5725A User Manual
Page 41

Introduction and Specifications
AC Current Specifications
1
1-25
Table 1-18. 5700A Series II AC Current Specifications: 99 % Confidence Level
5700A
99%
99 % Confidence Level
Range
Resolution
Frequency
Absolute Uncertainty
±
5 °C from calibration temperature
For fields strengths >0.4 V/m but
≤
3 V/m,
add 1 % of range.
Relative Uncertainty
±
1 °C
24 Hours
90 Days
180 Days
1 Year
24 Hours
90 Days
Hz
±
(ppm output + nA)
±
(ppm output + nA)
220
µ
A
1 nA
0
10 - 20
0
20 - 40
0
40 - 1 k
0
1k - 5 k
0
5k - 10 k
0
650 + 30
0
350 + 25
0
120 + 20
0
500 + 50
1500 + 100
0
700 + 30
0
380 + 25
0
140 + 20
0
600 + 50
1600 + 100
0
750 + 30
0
410 + 25
0
150 + 20
0
650 + 50
1700 + 100
00
800 + 30
00
420 + 25
00
160 + 20
00
700 + 50
0
1800 + 100
0
450 + 30
0
270 + 25
0
110 + 20
0
450 + 50
1400 + 100
0
500 + 30
0
300 + 25
0
120 + 20
0
500 + 50
1500 + 100
2.2 mA
10 nA
0
10 - 20
0
20 - 40
0
40 - 1 k
0
1k - 5 k
0
5k - 10 k
0
650 + 50
0
350 + 40
0
120 + 40
0
500 + 500
1500 + 1000
0
700 + 50
0
380 + 40
0
140 + 40
0
600 + 500
1600 + 1000
0
750 + 50
0
410 + 40
0
150 + 40
0
650 + 500
1700 + 1000
00
800 + 50
0 0
420 + 40
00
160 + 40
00
700 + 500
0
1800 + 1000
0
450 + 50
0
270 + 40
0
110 + 40
0
450 + 500
1400 + 1000
0
500 + 50
0
300 + 40
0
120 + 40
0
500 + 500
1500 + 1000
22 mA
100 nA
0
10 - 20
0
20 - 40
0
40 - 1 k
0
1k - 5 k
0
5k - 10 k
0
650 + 500
0
350 + 400
0
120 + 400
0
500 + 5000
1500 + 10,000
0
700 + 500
0
380 + 400
0
140 + 400
0
600 + 5000
1600 + 10,000
0
750 + 500
0
410 + 400
0
150 + 400
0
650 + 5000
1700 + 10,000
00
800 + 500
00
420 + 400
00
160 + 400
00
700 + 5000
0
1800 + 10,000
0
450 + 500
0
270 + 400
0
110 + 400
0
450 + 5000
1400 + 10,000
0
500 + 500
0
300 + 400
0
120 + 400
0
500 + 5000
1500 +
10,000
Hz
±
(ppm output +
µ
A)
±
(ppm output +
µ
A)
220 mA
1
µ
A
0
10 - 20
0
20 - 40
0
40 - 1 k
0
1k - 5 k
0
5k - 10 k
0
650 + 5
0
350 + 4
0
120 + 4
0
500 + 50
1500 + 100
0
700 + 5
0
380 + 4
0
150 + 4
0
600 + 50
1600 + 100
0
750 + 5
0
410 + 4
0
170 + 4
0
650 + 50
1700 + 100
00
800 + 5
00
420 + 4
00
180 + 4
0
700 + 50
0
1800 + 100
0
450 + 5
0
280 + 4
0
110 + 4
0
450 + 50
1400 + 100
0
500 + 5
0
300 + 4
0
130 + 4
0
500 + 50
1500 + 100
2.2 A
10
µ
A
0
20 - 1 k
0
1 k - 5 k
0
5 k - 10 k
0
600 + 40
0
700 + 100
8000 + 200
0
650 + 40
0
750 + 100
9000 + 200
0
700 + 40
0
800 + 100
9500 + 200
0
750 + 40
00
850 + 100
10,000 + 200
0
600 + 40
0
650 + 100
7500 + 200
0
650 + 40
0
750 + 100
8500 + 200
5725A Amplifier:
11 A
100
µ
A
0
40 - 1 k
0
1 k - 5 k
0
5 k - 10 k
0
370 + 170
0
800 + 380
3000 + 750
400 + 170
850 + 380
3300 + 750
0
440 + 170
0
900 + 380
3500 + 750
00
460 + 170
00
950 + 380
0
3600 + 750
0
300 + 170
0
700 + 380
2800 + 750
0
330 + 170
0
800 + 380
3200 + 750
Note: Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220
µ
A and 2.2 mA
ranges are increased by 1.3 x plus 2
µ
A when supplied through 5725A terminals. Specifications are otherwise
identical for all output locations.