Fluke 5725A User Manual
Page 134

5725A
Instruction Manual
7-4
Table 7-1. 5725A Fault Codes
FAULT
NO.
MESSAGE
EXPLANATION
201
5725 Self-Test ROM Failure
Requires service
202
5725 Self-Test RAM Failure
Requires service
203
5725 Self-Test EEPROM Failure
Requires service
204
5725 Self-Test Data Bus Failure
Requires service
205
5725 Self-Test CLAMPS Circuit Failure
Requires service
206
5725 Self-Test HVCLR Circuit Failure
Requires service
207
5725 Self-Test DAC Failure
Requires service
208
5725 Self-Test Watchdog Timer Failure
Requires service
209
5725 Current Heatsink Too Hot
Requires service
210
Output Tripped To Standby
Current heat sink too hot in operate
211
5725 Current Compliance Voltage Too
High
Internal fault, requires service
212
5725 Current Compliance Voltage Too
High
Most likely an external fault
213
5725 +400V Supply Did Not Shut Off
Requires service
214
5725 -400V Supply Did Not Shut Off
Requires service
215
5725 Voltage Heatsink Too Hot
Requires service
216
5725 Voltage Heatsink Too Hot
Voltage heat sink too hot in operate
217
5725 +400V Supply Too Small
Requires service, or improper line voltage
selected
218
5725 +400V Supply Too Large
Requires service, or improper line voltage
selected
219
5725 -400V Supply Too Large
Requires service, or improper line voltage
selected
220
5725 -400V Supply Too Small
Requires service, or improper line voltage
selected
221
5725 +400V Supply Current Too High
Internal fault, requires service
222
Output Tripped To Standby
+400V supply too high in operate, most likely an
external fault
223
5725 -400V Supply Current Too High
Internal fault, requires service
224
Output Tripped To Standby
-400V supply magnitude too high in operate,
most likely an external fault
225
5725 Fan Not Working
Requires service
226
5725 CLAMPS Fault
Internal fault, requires service
227
Output Tripped To Standby
CLAMPS circuit activated most likely because of
external transient condition