Fluke 5725A User Manual
Page 104

5725A
Instruction Manual
4-34
latch up by eliminating positive feedback resulting from positive output voltages from
U100. Op amp U105 is simply a low-noise inverting amplifier that scales the sensor
output to 10 dB below the 5700A Oscillator output level. The output of the calibration
circuit, AC/AC, goes to the 5700A during calibration through the analog monitor
section.
4-56. Analog Monitor Section
The 16-channel multiplexer, U151, takes a variety of inputs from all analog assemblies
and provides a means of switching them to either a comparator, U154A, or to the B-RCL
line via K152. During normal operation, the multiplexer output is routed exclusively to
the comparator. Calibration requires routing the multiplexer output to the B-RCL line,
which is routed back to the 5700A. The B-RCL line is measured by the precision adc
circuit on the 5700A DAC assembly. All inputs to the multiplexer are scaled so that
inputs are within its
±
15V power supply range. Table 4-4 shows the signal name of the
multiplexer input, its description, origin, and purpose.
The state of the multiplexer is controlled by the microprocessor via data latched into
U152.
Table 4-4. Signals Monitored by the Analog Monitor Section
STANDBY
VOLTAGE
LIMITS(1)
SIGNAL
NAME
PHYSICAL
PARAMETER
SOURCE
PURPOSE
-0.125V
+400V IM
Supply Current
Power Supply
Monitoring
+0.125V
-400V IM
Supply Current
Power Supply
Monitoring
0.26V
+400V M
Supply Voltage
Power Supply
Monitoring
-0.26V
-400V M
Supply Voltage
Power Supply
Monitoring
VCOM
System Supply Common
Power Supply
Monitoring
HVCOM M
High Voltage Supply
Common
Power Supply
Monitoring
1.7V
HVTEMP M
Voltage Amp Heat Sink
Temp.
High Voltage Amplifier
Monitoring
2.0V
ITEMP M
Current Amp Heat Sink
Temp.
Current Amplifier
Monitoring
-2.2V
ICOMPL M
Current Output Stage
Drive Level
Current Amplifier
Monitoring
I CAL/I FB
Shunt Voltage
Current Amplifier
Calibration, AC I
Operation
AC/AC
Calibrator Output
High Voltage Sense
Calibration
HVSENSE
Sense Amp Output
High Voltage Sense
Calibration
DAC OUT
DAC Output
High Voltage Sense
Monitoring
0.08, 2.0V
FAN M
Fan is On or Off
Power Supply
Monitoring
Note 1: For single limit values, the limit is equal to, or smaller in magnitude to, the value shown.