Fluke 5725A User Manual
Page 24

5725A
Instruction Manual
1-8
Table 1-2. 5700A Series II DC Voltage Specifications: 99 % and 95 % Confidence Levels
5700A
99%
99 % Confidence Level
Range
Resolution
Absolute Uncertainty
±
5 °C from calibration temperature
For fields strengths >1 V/m but
≤
3 V/m,
add 0.01 % of range.
Relative Uncertainty
±
1 °C
24 Hours
90 Days
180 Days
1 Year
24 Hours
90 Days
±
(ppm output +
µ
V)
±
(ppm output +
µ
V)
220 mV
10 nV
6.5 + .75
7 + .75
0
8 + .75
0
9 + .8
2.5 + .5
0
4 + .5
2.2 V
100 nV
3.5 + 1.2
6 + 1.2
0
7 + 1.2
0
8 + 1.2
2.5 + 1.2
0
4 + 1.2
11 V
1
µ
V
3.5 + 3
5 + 4
0
7 + 4
0
8 + 4
1.5 + 3
3.5 + 4
22 V
1
µ
V
3.5 + 6
5 + 8
0
7 + 8
0
8 + 8
1.5 + 6
3.5 + 8
220 V
10
µ
V
0
5 + 100
6 + 100
0
8 + 100
0
9 + 100
2.5 + 100
0
4 + 100
1100 V
100
µ
V
0
7 + 600
8 + 600
10 + 600
11 + 600
0
3 + 600
4.5 + 600
5700A
95%
95 % Confidence Level
Range
Resolution
Absolute Uncertainty
±
5 °C from calibration temperature
For fields strengths >1 V/m but
≤
3 V/m,
add 0.01 % of range.
Relative Uncertainty
±
1 °C
24 Hours
90 Days
180 Days
1 Year
24 Hours
90 Days
±
(ppm output +
µ
V)
±
(ppm output +
µ
V)
220 mV
10 nV
5.5 + 0.6
6 + 0.6
7 + 0.6
8 + 0.6
0
2 + 0.4
3.5 + 0.4
2.2 V
100 nV
3.5 + 1
5 + 1
6 + 1
7 + 1
0
2 + 1
3.5 + 1
11 V
1 mV
0
3 + 3.5
4 + 3.5
6 + 3.5
7 + 3.5
1.2 + 3
0
3 + 3.5
22 V
1 mV
0
3 + 6.5
4 + 6.5
6 + 6.5
7 + 6.5
1.2 + 6
0
3 + 7
220 V
10 mV
0
4 + 80
5 + 80
7 + 80
8 + 80
0
2 + 80
3.5 + 80
1100 V
100 mV
0
6 + 500
7 + 500
8 + 500
9 + 500
2.4 + 500
0
4 + 500