5725a – Fluke 5725A User Manual
Page 30

5725A
Instruction Manual
1-14
Table 1-8. AC Voltage Secondary Performance Specifications and Operating Characteristics
Stability
±
1 °C [Note 1]
Temperature Coefficient
Maximum
Distortion
Range
Frequency
24 Hours
10°-40 °C
0°-10 °C and
40°-50 °C
Output Impedance
Bandwidth
10 Hz-10 MHz
Hz
± µ
V
±µ
V/°C
Ω
±
(% output +
µ
V)
2.2 mV
00
10 - 20
00
20 - 40
00
40 - 20 k
0
20 k - 50 k
0
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
5
5
2
2
3
3
5
5
0.05
0.05
0.05
0.1
0.2
0.3
0.4
0.5
0.05
0.05
0.05
0.1
0.2
0.3
0.4
0.5
50
0
0.05 + 10
0.035 + 10
0.035 + 10
0.035 + 10
0.035 + 10
00
0.3 + 30
00
0.3 + 30
000
1 + 30
22 mV
00
10 - 20
00
20 - 40
00
40 - 20 k
0
20 k - 50 k
0
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
5
5
2
2
3
5
10
15
0.2
0.2
0.2
0.4
0.5
0.6
1
1
0.3
0.3
0.3
0.5
0.5
0.6
1
1
50
0
0.05 + 11
0.035 + 11
0.035 + 11
0.035 + 11
0.035 + 11
00
0.3 + 30
00
0.3 + 30
000
1 + 30
±
(ppm output +
µ
V)
±
(ppm output
µ
V)/°C
220 mV
00
10 - 20
00
20 - 40
00
40 - 20 k
0
20 k - 50 k
0
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 20
0
80 + 15
0
12 + 2
0
10 + 2
0
10 + 2
0
20 + 4
100 + 10
200 + 20
00
2 + 1
00
2 + 1
00
2 + 1
0
15 + 2
0
15 + 4
0
80 + 5
0
80 + 5
0
80 + 5
0
2 + 1
0
2 + 1
0
2 + 1
15 + 2
15 + 4
80 + 5
80 + 5
80 + 5
50
0
0.05 + 16
0.035 + 16
0.035 + 16
0.035 + 16
0.035 + 16
00
0.3 + 30
00
0.3 + 30
000
1 + 30
Load Regulation
±
(ppm output+
µ
V)
2.2 V
00
10 - 20
00
20 - 40
00
40 - 20 k
0
20 k - 50 k
0
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 20
0
80 + 15
0
12 + 4
0
15 + 5
0
15 + 5
0
30 + 10
0
70 + 20
150 + 50
0
50 + 10
0
15 + 5
00
2 + 1
0
10 + 2
0
10 + 4
0
80 + 15
0
80 + 40
0
80 + 100
50 + 10
15 + 5
0
5 + 2
15 + 4
20 + 4
80 + 15
80 + 40
80 + 100
00
10 + 2
00
10 + 2
00
10 + 4
00
30 + 10
0
120 + 16
0
300 ppm
0
600 ppm
0
1200 ppm
0
0.05 + 80
0.035 + 80
0.035 + 80
0.035 + 80
0.035 + 80
00
0.3 + 110
00
0.3 + 110
000
1 + 110
22 V
00
10 - 20
00
20 - 40
00
40 - 20 k
0
20 k - 50 k
0
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 20
0
80 + 15
0
12 + 8
0
15 + 10
0
15 + 10
0
30 + 15
0
70 + 100
150 + 100
0
50 + 100
0
15 + 30
00
2 + 10
0
10 + 20
0
10 + 40
0
80 + 150
0
80 + 300
0
80 + 500
50 + 100
15 + 40
0
4 + 15
20 + 20
20 + 40
80 + 150
80 + 300
80 + 500
00
10 + 20
00
10 + 20
00
10 + 30
00
30 + 50
00
80 + 80
0
100 + 700
0
200 + 1100
0
600 + 3000
0
0.05 + 700
0.035 + 700
0.035 + 700
0.035 + 700
0.035 + 700
00
0.3 + 800
00
0.3 + 800
000
2 + 800
220 V
00
10 - 20
00
20 - 40
00
40 - 20 k
0
20 k - 50 k
0
50 k - 100 k
100 k - 300 k
300 k - 500 k
500 k - 1 M
150 + 200
0
80 + 150
0
12 + 80
0
15 + 100
0
15 + 100
0
30 + 400
100 + 10,000
200 + 20,000
0
50 + 1000
0
15 + 300
00
2 + 80
0
10 + 100
0
10 + 500
0
80 + 600
0
80 + 800
0
80 + 1000
50 + 1000
15 + 300
0
4 + 80
20 + 100
20 + 500
80 + 600
80 + 800
80 + 1000
00
10 + 200
00
10 + 200
00
10 + 300
00
30 + .600
00
80 + 3,000
0
250 + 25,000
0
500 + 50,000
1000 + 110,000
0
0.05 + 10,000
0
0.05 + 10,000
0
0.05 + 10,000
0
0.05 + 10,000
00
0.1 + 13,000
00
1.5 + 50,000
00
1.5 + 50,000
00
3.5 + 100,000
±
(ppm output + mV)
±
(ppm output)/°C
±
(% output)
1100 V
00
15 - 50
00
50 - 1 k
150 + 0.5
0
20 + 0.5
50
2
50
5
00
10 + 2
00
10 + 1
0.15
0.07
Note: 1. Stability specifications are included in Absolute Uncertainty values for the primary specifications.