Troubleshooting
Self Diagnostics
7
7-5
Table 7-1. 5725A Fault Codes (cont)
FAULT
NO.
MESSAGE
EXPLANATION
228
5725 Software TRAP
229
5725 Cable Was Off
230
5725 RESET (power-up or watchdog timer)
231
5725 Guard-Crossing Timeout
232
5725 Illegal/Unexecutable Command
233
5725 Non-Maskable Interrupt Occurred
Either internal fault or external transient condition
234
5725 HVCLEAR Circuit Activated
Internal fault, requires service
235
Output Tripped To Standby
HVCLEAR circuit activated most likely because of
external transient condition
841
No 5725 Available For Selected Output
850
Can’t Store, 5725 Switch In NORMAL
851
Can’t Format, 5725 Switch In NORMAL
3202
5725 ACV Sense Amp Fault
Requires service
3203
5725 ACV Standby 5725 Fault
Requires service
3204
5725 ACV Operate Wouldn’t Converge
Requires service
3205
5725 ACV Operate 5725 Fault
Requires service
3206
5725 ACV Cal Sensor Test Died
Requires service
3207
5725 ACV Cal Sensor Fault
Requires service
3208
5725 Current Path To Shunt Open
Requires service
3209
5725 Shunt Sense Open
Requires service
3210
5725 Shunt Measurement Out Of
Tolerance
Requires service
3211
5725 Current Amplifier Offset Too Large
Requires service
3212
5725 Current Drive Path Open
Requires service
3213
5725 Current Error Amplifier Failure
Requires service
6.
To test the current function, press the "Current" softkey. A sequence of self tests is
run. A brief description of each test follows:
a.
5725A Current Shunt: Checks the current cabling and the calibration amplifier
on the Current Amplifier assembly (A2).
b.
5725A Current Zero: Grounds the input and shorts the output of the Current
Amplifier, and verifies that the output is close to 0A.
c.
5725A Current Gain: Shorts the output of the Current Amplifier, applies a
voltage input, and verifies the output current. This confirms integrity of the error
amplifier circuit in the Current Amplifier assembly.
d.
5725A Current Monitor: Runs the analog monitor circuit on the High Voltage
Sense Assembly (A6) in the current function and verifies that it operates
correctly.