5725a – Fluke 5725A User Manual
Page 40

5725A
Instruction Manual
1-24
Table 1-17. 5720A Series II AC Current Specifications: 95 % Confidence Level
5720A
95%
95 % Confidence Level
Range
Resolution
Frequency
Absolute Uncertainty
±
5 °C from calibration temperature
For fields strengths >0.4 V/m but
≤
3 V/m,
add 1 % of range.
Relative Uncertainty
±
1 °C
24 Hours
90 Days
180 Days
1 Year
24 Hours
90 Days
Hz
±
(ppm output + nA)
±
(ppm output + nA)
220
µ
A
1 nA
0
10 - 20
0
20 - 40
0
40 - 1 k
0
1k - 5 k
0
5k - 10 k
0
210 + 16
0
130 + 10
0
100 + 8
0
240 + 12
0
800 + 65
0
230 + 16
0
140 + 10
0
110 + 8
0
250 + 12
0
900 + 65
0
240 + 16
0
150 + 10
0
115 + 8
0
270 + 12
1000 + 65
0
250 + 16
0
160 + 10
0
120 + 8
0
280 + 12
1100 + 65
0
210 + 16
0
110 + 10
00
80 + 8
0
200 + 12
0
700 + 65
0
230 + 16
0
130 + 10
00
90 + 8
0
230 + 12
0
800 + 65
2.2 mA
10 nA
0
10 - 20
0
20 - 40
0
40 - 1 k
0
1k - 5 k
0
5k - 10 k
0
210 + 40
0
140 + 35
0
100 + 35
0
170 + 110
0
800 + 650
0
230 + 40
0
140 + 35
0
110 + 35
0
180 + 110
0
900 + 650
0
240 + 40
0
150 + 35
0
115 + 35
0
190 + 110
1000 + 650
0
250 + 40
0
160 + 35
0
120 + 35
0
200 + 110
1100 + 650
0
210 + 40
0
110 + 35
00
80 + 35
0
200 + 110
0
700 + 650
0
230 + 40
0
130 + 35
00
90 + 35
0
230 + 110
0
800 + 650
22 mA
100 nA
0
10 - 20
0
20 - 40
0
40 - 1 k
0
1k - 5 k
0
5k - 10 k
0
210 + 400
0
130 + 350
0
100 + 350
0
170 + 550
0
800 + 5000
0
230 + 400
0
140 + 350
0
110 + 350
0
180 + 550
0
900 + 5000
0
240 + 400
0
150 + 350
0
115 + 350
0
190 + 550
1000 + 5000
0
250 + 400
0
160 + 350
0
120 + 350
0
200 + 550
1100 + 5000
0
210 + 400
0
110 + 350
00
80 + 350
0
200 + 550
0
700 + 5000
0
230 + 400
0
130 + 350
00
90 + 350
0
230 + 550
0
800 + 5000
Hz
±
(ppm output +
µ
A)
±
(ppm output +
µ
A)
220 mA
1
µ
A
0
10 - 20
0
20 - 40
0
40 - 1 k
0
1k - 5 k
0
5k - 10 k
0
210 + 4
0
130 + 3.5
0
100 + 2.5
0
170 + 3.5
0
800 + 10
0
230 + 4
0
140 + 3.5
0
110 + 2.5
0
180 + 3.5
0
900 + 10
0
240 + 4
0
150 + 3.5
0
115 + 2.5
0
190 + 3.5
1000 + 10
0
250 + 4
0
160 + 3.5
0
120 + 2.5
0
200 + 3.5
1100 + 10
0
210 + 4
0
110 + 3.5
00
80 + 2.5
0
200 + 3.5
0
700 + 10
0
230 + 4
0
130 + 3.5
0
90 + 2.5
0
230 + 3.5
0
800 + 10
2.2 A
10
µ
A
0
20 - 1 k
0
1 k - 5 k
0
5 k - 10 k
0
230 + 35
0
350 + 80
5000 + 160
0
240 + 35
0
390 + 80
6000 + 160
0
250 + 35
0
420 + 80
6500 + 160
0
260 + 35
0
450 + 80
7000 + 160
0
250 + 35
0
400 + 80
5000 + 160
0
300 + 35
0
440 + 80
6000 + 160
5725A Amplifier:
11 A
100
µ
A
0
40 - 1 k
0
1 k - 5 k
0
5 k - 10 k
0
370 + 170
0
800 + 380
3000 + 750
0
400 + 170
0
850 + 380
3300 + 750
0
440 + 170
0
900 + 380
3500 + 750
0
460 + 170
0
950 + 380
3600 + 750
0
300 + 170
0
700 + 380
2800 + 750
0
330 + 170
0
800 + 380
3200 + 750
Note: Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220
µ
A and
2.2 mA ranges are increased by 1.3 x plus 2
µ
A when supplied through 5725A terminals. Specifications are
otherwise identical for all output locations.