Error Messages
Error
Description
Err 0
Power on zero error. See parameter 20.
Err 1
Keyboard error. Occurs during power up while a key on the keyboard is
being pressed or one of the remote switch inputs is activated.
Err 2
Restart Trap. The microprocessor accessed nonexistent memory location.
Usual cause is A/C power “glitch” or static electric discharge.
Err 3
Watchdog Time-out. The weight display has not been updated within the
watchdog time-out period. Usual cause is A/C power glitch or static
electric discharge.
Err 4
Battery Error. Battery voltage level was below 2 V when the DT200 was
powered off.
Err 5
Earom memory error. A checksum error has occurred when reading the
configuration earom memory. Check calibration parameters and rewrite the
earom memory. If error continues, replace earom (U13)
Err 5.1
Earom Time-out, unable to write to the earom memory.
Err 6
Ram memory error. A ram memory error has occurred. Replace U8 if error
continues.
Err 7
Err 8
Err 9
Count-by error. The entered value is outside the scale count-by set during
calibration.
Err 10
Err 11
Err 12
Err 13
Printer busy error. The busy signal (TB4 pin 14) is active (+5v TTL hi).
Err 14
Page Format line length error. The page format output exceeded the
maximum line length.
Err 15
The function selected is locked. Switch 2 on the DT200’s circuit board
locks and unlocks configuration parameters.
Err 16
Scale Basic Stack Overflow.
Err 18
Referenced an invalid function instruction.
Err 21
ID not found.
Err 22
Scale Basic Prompt error.
Err 23
ID memory full.
Error
OL
Scale is over-loaded (above scale capacity).