Diagnostic Tests
38
4. Use the UNITS key (
↓) or the PRINT key (↑) to scroll through the entry numbers.
Press the CLEAR key to exit from the Function Editor.
Function 73 Initialize Function Memory
The initialize function erases function memory and setpoint monitor memory, and
initializes all parameters with password level 1 or 2.
Function 73 prompts “rs Fn” . Press the Enter key to initialize only or enter a number
below to activate a built in program.
1. Fill to Setpoint.
2. Checkweigh: Over/Between/Under
3. Weigh-in / Weigh-out
4. Axle Weigh – Short Scale
5. Axle Weigh – Long Scale
6. ID Tare and Total
7. Peak Hold
Function 74 Configure Timers
The display prompts “tr” when function 74 is activated. Enter the timer number to be
modified (1 to 5) then press the ENTER key. The first parameter (L) is the time length in
tenths of a second. The second parameter (Fn) is the function number to be executed
when the timer times out.
Parameter 76 Debug mode for Event monitors and Scale Basic
The Debug mode is used to ‘debug’ event monitors and scale basic functions. Enter a 1
to activate debug mode, enter a 0 to disable debug mode. When debug mode is enabled
Scale Basic functions ‘single step’. The subroutine stack level and the Function number
are displayed when entering a function. Example: “0 F12”. Press the ENTER key to
begin execution of the function. The instruction address and instruction code are
displayed. Example: “21 255”. Continue to press the ENTER key to step thru the
program. NOTE: the first step displayed for all scale basic functions is a GOTO (244).
Function 78 Initialize ID memory
Use function 78 to initialize id memory after it has been corrupted. The ID memory
becomes corrupted when power is lost (when battery backup is disabled) or when the
battery is allowed to discharge below 2 volts.
Parameters 91, 92, 93: Serial Port 3
Serial Communications Port 3 receives weight data from a weight indicator. The mode,
baud rate, and format of com. port 3 must be the same as the weight indicator’s settings.