Industrial Data Systems IDS DT200 User Manual
Page 30

DT 200 Users Manual version 1.E
Set Operation Parameters
26
2 - CONDEC:
Polarity:
space (positive weight), - (negative weight)
Weight:
7 characters, zero blanked, with decimal point
Units:
L (lb), K (kg), T (ton), G (gram), O (oz)
Mode:
G (gross mode), N (net mode)
Status:
Terminate:
lf, cr (line feed, carriage return)
3 - WEIGHTRONIX:
Mode1:
space (gross mode), 0 (zero if net mode)
Mode2:
G (gross mode), N (net mode)
Polarity:
+ (positive weight), - (negative weight)
Weight:
6 characters, NOT zero blanked, with decimal point
Units:
lb, kg, oz, g
Terminate:
lf, cr (line feed, carriage return)
4 - OHAS:
Polarity:
space (positive weight), - (negative weight)
Weight:
7 digits, zero blanked, with decimal point
Units:
lb, kg, oz, g
Mode:
G (gross mode), N (net mode)
Status:
Terminate:
lf, cr (line feed, carriage return)
5 - Print label 85: transmits the contents of parameter 85 (print label 85)
Data entered into a print label can include register data:
[253] [reg no] buffer register data, minimum size (zero blanked no spaces)
[254] [reg no] buffer data, 7 character field, zero blanked, right justified.
Printer Parameters
The DT200 provides a universal printer interface for various printers. The printer
interface allows you to select and customize the data output for serial or parallel printers.
The serial interface option uses Serial Port 1 (TB1) for connection to the printer. The
parallel interface uses the TTL I/O Port (TB4) for connection to the printer. Refer to the
“HARDWARE INSTALLATION AND WIRING “ section for information on connector
pin designations. NOTE: The parallel interface may not be available if the DT200 is
configured to use setpoint outputs.
Parameter 29 Printer Port Select (2 [Serial Port 1])
Enter a number from the following list to select a printer option:
1 Parallel Printer interface.