beautypg.com

Appendix c – HP E1368A User Manual

Page 80

background image

Appendix C

HP E1368A/69A/70A Error Messages

Table C-1 lists the error messages associated with the Microwave Switch
module programmed by SCPI. See the appropriate mainframe manual for a
complete list of error messages.

Table C-1. Microwave Switch Error Messages

No.

Title

Potential Cause(s)

-211

Trigger ignored

Trigger received when scan not enabled. Trigger received after scan

complete. Trigger too fast.

-213

INIT ignored

Attempted to execute an INIT command when a scan is already in
progress.

-224

Illegal parameter value

Attempting to execute a command with a parameter not applicable to
the command.

1500 External trigger source

already allocated

Assigning an external trigger source to a switchbox when the trigger
source has already been assigned to another switchbox.

2000 Invalid card number

Addressing a module (card) in a switchbox that is not part of the
switchbox.

2001 Invalid channel number

Attempting to address a channel of module in a switchbox that is not
supported by the module (e.g., channel 99 of a multiplexer module).

2006 Command not supported

on this card

Sending a command to a module (card) in a switchbox that is
unsupported by the module.

2008 Scan list not initialized

Executing a scan without the INIT command.

2009 Too many channels in

channel list

Attempting to address more channels than available in the switchbox.

2010 Scan mode not supported

on this card

Sending a command to a module (card) in a switchbox that is
unsupported by the module

2012 Invalid Channel Range

Invalid channel(s) specified in SCAN <

channel_list

> command.

Attempting to begin scanning when no valid channel list is defined.

2600 Function not supported on

this card

Sending a command to a module (card) in a switchbox that is not
supported by the module or switchbox.

2601 Channel list required

Sending a command requiring a channel list without the channel list.

Appendix C

HP E1368A/69A/70A Error Messages 79

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com