Initiate, Continuous – HP E1368A User Manual
Page 56
INITiate
The
INITiate
subsystem selects continuous scanning cycles and starts the
scanning cycle.
Subsystem Syntax
INITiate
:CONTinuous <
mode
>
:CONTinuous?
[:IMMediate]
:CONTinuous
INITiate:CONTinuous <
mode
>
enables or disables continuous scanning
cycles for the switchbox.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Value
mode
boolean
0 | 1 | ON | OFF
OFF | 0
Comments
•
Continuous Scanning Operation: Continuous scanning is enabled
with the
INITiate:CONTinuous ON
or
INITiate:CONTinuous 1
command. Sending the
INITiate[:IMMediate]
command closes the
first channel in the channel list. Each trigger from a trigger source
selected by the
TRIGger:SOURce
command advances the scan
through the channel list. A trigger at the end of the channel list
closes the first channel in the list and the scan cycle repeats.
•
Non-Continuous Scanning Operation: Non-continuous scanning is
enabled with the
INITiate:CONTinuous OFF
or
INITiate:CONTinuous
0
command. Sending the
INITiate[:IMMediate]
command closes the
first channel in the channel list. Each trigger from a trigger source
selected by the
TRIGger:SOURce
command advances the scan
through the channel list. At the end of the scanning cycle, the last
channel in the channel list is closed.
•
Stopping Continuous Scans: See the
ABORt
command.
•
Related Commands:
ABORt
,
ARM:COUNt
,
TRIGger
,
TRIGger:SOURce
Example
Enabling Continuous Scans
This example enables continuous scanning of channels 00 through 02 of a
single-module switchbox. Since
TRIGger:SOURce IMMediate
(default) is
set, use an interface clear command (such as CLEAR) to stop the scan.
INIT:CONT ON
Enables continuous scanning.
SCAN (@100:102)
Defines channel list.
INIT
Start scan cycle, close channel 00.
Chapter 4
HP E1368A/69A/70A Command Reference 55
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com