Output, State – HP E1368A User Manual
Page 58
OUTPut
The
OUTPut
subsystem enables or disables the "Trig Out" port of the HP
E1300/E1301 Mainframe or on the HP Command Module.
Subsystem Syntax
OUTPut
[:STATe] <
state
>
[:STATe]?
[:STATe]
OUTPut[:STATe] <
state
>
enables/disables the "Trig Out" port on the rear
panel of the HP E1300A/E1301A Mainframe or on the HP Command
Module.
OUTPut[:STATe] ON | 1 enables the port and OUTPut[:STATe]
OFF | 0
disables the port.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Value
state
boolean
0 | 1 | ON | OFF
OFF | 0
Comments
•
Enabling "Trig Out" Port: When enabled, a pulse is output from
the "Trig Out" port after each switchbox channel is closed. If
disabled, a pulse is not output from the "Trig Out" port after channel
closures. The output pulse is a +5V negative-going pulse.
•
"Trig Out" Port Shared by Switchboxes: When enabled, the "Trig
Out" port may be pulsed by any switchbox each time a scanned
channel is closed. To disable the output for a specific switchbox,
send the
OUTPut[:STATe] OFF
or
OUTPut[:STATe] 0
command for
that switchbox.
•
Related Commands:
[ROUTE:]SCAN
,
TRIGger:SOURce
•
*RST Condition:
OUTPut[:STATe] OFF
(port disabled)
Example
Enabling "Trig Out" Port
OUTP:STAT ON
Enables "Trig Out" port to output
pulse after each scanned channel is
closed.
[:STATe]?
OUTPut[:STATe]?
queries the present state of the "Trig Out" port. The
command returns 1 if the port is enabled, or 0 if disabled.
Example
Query "Trig Out" Port Enable State
This example enables the "Trig Out" port and queries the enable state. The
OUTPut[:STATe]?
command returns 1 since the port is enabled.
OUTP:STAT ON
Enables "Trig Out" port for pulse
output.
OUTP:STAT?
Query port enable state.
Chapter 4
HP E1368A/69A/70A Command Reference 57
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com