SCPI Command Reference
This section describes the Standard Commands for Programmable
Instruments (SCPI) commands for the Microwave Switch modules.
Commands are listed alphabetically by subsystem and also within each
subsystem.
ABORt
The
ABORt
subsystem stops a scan in progress when the scan is enabled via
the interface, and the trigger modes are
TRIGger:SOURce BUS
or
TRIGger:SOURce HOLD
.
Subsystem Syntax
ABORt
Comments
•
ABORt
Operation invalidates the current channel list and sets
ARM:COUNt 1
(one scanning cycle per
INITiate
command), sets
INITiate CONTinuous OFF
(no continuous scanning cycles), and sets
TRIGger:SOURce IMMediate
(continuous internal triggering).
•
Stopping Scans Enabled from Interface: When a scan is enabled
from the interface, use an interface
CLEAR
command or the HP
E1301 front panel "Reset Instr" or "Clear Instr" key to stop the scan.
When the scan is enabled from the interface and the trigger source is
TRIGger:SOURce BUS
or
TRIGger:SOURce HOLD
, use
ABORt
or
the HP E1301 front panel "Reset Instr" or "Clear Instr" keys to stop
the scan.
•
Stopping Scans Enabled from Front Panel: When a scan is
enabled from the HP E1301 front panel, execute
*RST
over the
interface or the front panel "Reset Instr" or "Clear Instr" keys to stop
the scan.
•
Related Commands:
ARM
,
INITiate:CONTinuous
,
[ROUTe:]SCAN
,
TRIGger
Example
Stopping a Scan with
ABORt
This example stops a (continuous) scan in progress.
TRIG:SOUR BUS
BUS is trigger source.
INIT:CONT ON
Set continuous scanning.
SCAN (@100:102)
Scan channels 00-02.
INIT
Starts scanning cycle, close
channel 00.
ABOR
Aborts scan in progress.