beautypg.com

Count – HP E1368A User Manual

Page 53

background image

:COUNt?

ARM:COUNt? [MIN | MAX]

returns the current number of scanning cycles

set by

ARM:COUNt

. The current number of scans cycles is returned when

no

MIN

or

MAX

parameter is sent. With

MIN

or

MAX

passed as parameters,

MIN

returns 1 and

MAX

returns 32767.

Parameters

Parameter Name

Parameter Type

Range of Values

Value

MIN | MAX

numeric

MIN = 1, MAX = 32767

cycles

Comments

Related Commands:

INITiate:IMMediate

Example

Query Number of Scanning Cycles

This example sets a Microwave Switch for 10 scanning cycles of channels
00 through 02, and queries the number of scan cycles set. The

ARM:COUN?

command returns 10.

ARM:COUN 10

Set 10 scans per INIT command.

ARM COUN?

Query number of scanning cycles.

52 HP E1368A/69A/70A Command Reference

Chapter 4

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com