Display, Monitor:card – HP E1368A User Manual
Page 54
DISPlay
The
DISPlay
subsystem monitors the channel state of a selected module (or
card) in a switchbox. This subsystem only operates with mainframes that
have a display, such as the HP 75000 Series B Mainframe (HP Model
Number E1301A).
Subsystem Syntax
DISPlay
:MONitor
:CARD
<
number
| AUTO>
[:STATe] <
mode
>
[:STATe]?
:MONitor:CARD
DISPlay:MONitor:CARD <
number |
AUTO>
selects the module in a
switchbox to be monitored when the monitor mode is enabled. Use the
DISPlay:MONitor:STATe
command to enable and disable the monitor mode.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Value
number
| AUTO
numeric
1 - 99 or AUTO
AUTO
Comments
•
Selecting a Specific Module to be Monitored: Use the
DISPlay:MONitor:CARD
command to send the card number in a
switchbox to be monitored.
•
Selecting the Present Module to be Monitored: Use the
DISPlay:MONitor AUTO
command to select the last module
addressed by a switching command (e.g.,
[ROUTe:]CLOSe
).
•
*RST Condition:
DISPlay:MONitor:CARD AUTO
Example
Select Module #2 in a Switchbox for Monitoring
DISP:MON:CARD 2
Selects module #2 in a switchbox
to be monitored when the mode is
enabled.
Chapter 4
HP E1368A/69A/70A Command Reference 53
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com