:OPERation:ENABle
STATus:OPERation:ENABle <
number
>
enables the Standard Operation
Status Register to set a bit in the Status Byte Register. For the Microwave
Switch, when bit 8 in the Standard Operation Status Register is set to 1, bit
7 in the Status Byte Register is set to 1.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Value
number
numeric
0 - 32767
0
Comments
•
Summary Bit: If bit 8 is 1 in the Standard Operation Status Enable
Register, a transition of Event Register bit 8 to 1 will cause the
associated summary bit to be true.
•
Maximum Value Entered: The maximum decimal weighted value
used in the Microwave Switch module is 256 (bit 8 set true).
•
Setting Bit 7 of the Status Byte Register:
STATus:OPERation:ENABle 256
sets bit 7 of the Status Byte
Register to 1 after bit 8 of the Standard Operation Status Enable
Register is set to 1.
•
Related Commands:
[ROUTe:]SCAN
Example
Enable the Status Byte Register Bit 7
STAT:OPER:ENAB 256
Enables bit 7 of the Standard
Operation Status Enable Register,
which enables bit 7 of Status Byte
Register.
:OPERation:ENABle?
STATus:OPERation:ENABle?
returns the bit value of the Standard
Operation Status Enable Register.
Comments
•
Output Format: Returns a decimal weighted value from 0 to 65535
indicating which bits are set true.
•
Maximum Value Returned: The maximum decimal weighted value
used in the Microwave Switch module is 256 (bit 8 set true).
Example
Query the Standard Operation Status Enable Register
STAT:OPER:ENAB?
Queries the Standard Operation
Status Enable Register.
Chapter 4
HP E1368A/69A/70A Command Reference 63