beautypg.com

Operation:enable – HP E1368A User Manual

Page 64

background image

:OPERation:ENABle

STATus:OPERation:ENABle <

number

>

enables the Standard Operation

Status Register to set a bit in the Status Byte Register. For the Microwave
Switch, when bit 8 in the Standard Operation Status Register is set to 1, bit
7 in the Status Byte Register is set to 1.

Parameters

Parameter Name

Parameter Type

Range of Values

Default Value

number

numeric

0 - 32767

0

Comments

Summary Bit: If bit 8 is 1 in the Standard Operation Status Enable
Register, a transition of Event Register bit 8 to 1 will cause the
associated summary bit to be true.

Maximum Value Entered: The maximum decimal weighted value
used in the Microwave Switch module is 256 (bit 8 set true).

Setting Bit 7 of the Status Byte Register:

STATus:OPERation:ENABle 256

sets bit 7 of the Status Byte

Register to 1 after bit 8 of the Standard Operation Status Enable
Register is set to 1.

Related Commands:

[ROUTe:]SCAN

Example

Enable the Status Byte Register Bit 7

STAT:OPER:ENAB 256

Enables bit 7 of the Standard
Operation Status Enable Register,
which enables bit 7 of Status Byte
Register.

:OPERation:ENABle?

STATus:OPERation:ENABle?

returns the bit value of the Standard

Operation Status Enable Register.

Comments

Output Format: Returns a decimal weighted value from 0 to 65535
indicating which bits are set true.

Maximum Value Returned: The maximum decimal weighted value
used in the Microwave Switch module is 256 (bit 8 set true).

Example

Query the Standard Operation Status Enable Register

STAT:OPER:ENAB?

Queries the Standard Operation
Status Enable Register.

Chapter 4

HP E1368A/69A/70A Command Reference 63

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com