Reset conditions – HP E1368A User Manual
Page 31
Reset Conditions
When the Microwave Switch is switched on or
*RST
(reset), all channels
are set to open (refer to Figure 2-6), and the current channel list for scanning
is invalidated. Table 3-2 lists the parameters and default values for the
functions described in Chapter 3.
Table 3-2. *RST (Reset) Default Conditions and Values
Parameter
Default
Description
ARM:COUNt
1
Number of scanning cycles is one.
TRIGger:SOURce
IMM
Will advance scanning cycles automatically.
INITiate:CONTinuous
OFF
Number of scanning cycles is determined by
ARM:COUNt.
OUTPut:STATe
OFF
"Trig Out" port on mainframe is disabled.
SCAN:MODE
NONE Not used by the Microwave Switch modules.
SCAN:PORT
NONE Not used by the Microwave Switch modules.
Caution
When the Microwave Switch is powered up, signals connected
to the normally closed (N/C) port will also be connected to the
output port (refer to Figure 2-6).
Note
The Microwave Switch modules do not support 4-wire resistance scanning
(FRES).
30 Using the HP E1368A/69A/70A Modules
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com