Connecting Field
Wiring
Figure 2-8 shows the internal switch diagram for the various HP
single-pole, multi-throw switches. All switches are shown in the "OPEN"
position. To minimize loss at high frequencies, use the following guidelines
when making connections.
•
For frequencies up to 18 GHz, use a good quality flexible type cable
and SMA connectors.
•
For frequencies >18 GHz, use semi-rigid type cable and APC 3.5
connectors.
Figure 2-9 shows the internal schematic of a single attenuator section.
Figure 2-9. Attenuator
Figure 2-8. Internal Switch Diagrams
28 Configuring the HP E1368A/69A/70A Modules
Chapter 2