Ieee 488.2 common commands – HP E1368A User Manual
Page 71
IEEE 488.2 Common Commands
The following table lists the IEEE 488.2 Common (*) Commands that can
be executed by the Microwave Switch modules. The operation of some of
these commands is described in Chapter 3 of this manual. For more
information on Common Commands, refer to the HP 75000 Series B
Mainframe User’s Manual (HP Model Number E1300/E1301), appropriate
HP Command Module User’s Manual, or the ANSI/IEEE Standard
488.2-1987.
Command
Title
Description
*IDN?
Identification query
Returns identification string of the switchbox.
*RST
Reset
Opens all channels, and invalidates current
channel list
for scanning.
Sets ARM:COUN 1, TRIG:SOUR IMM, and INIT:CONT OFF.
*TST?
Self-Test
Always returns 0.
*OPC
Operation Complete
Sets the request for OPC flag when all pending operations have
completed. Also sets OPC bit in the Standard Event Register.
*OPC?
Operation Complete Query
Returns a 1 to the output queue when all pending operations have
completed. Used to synchronize between multiple instruments.
*WAI
Wait to Complete
Halts execution of commands and queries until the "No Operation
Pending" message is true.
*CLS
Clear status
Clears all Event Registers, the Request for OPC flag, and all queues
(except output queue).
*ESE<
mask
>
Event status enable
Used to set the bits in the Event Status Enable Register.
*ESE?
Event status enable query
Queries the current contents in the Event Status Enable Register.
*ESR?
Event status register query
Queries and clears the contents in the Standard Event Status Register.
*SRE<
mask
>
Service request enable
Used to set the Service Request Enable Register bits, and corresponding
Serial Poll Status Byte Register bits, to generate a service request.
*SRE?
Service request enable query
Queries the current contents in the Service Request Enable Register.
*STB?
Read status byte query
Queries the current contents in the Status Byte Register.
*TRG
Trigger
Triggers the switchbox to advance the scan when scan is enabled and
trigger source is TRIGger:SOURce BUS.
*RCL<
n
>
Recall saved state
Recalls previously stored Microwave Switch module configuration.
<
n
> (0 to 9) is the location in memory where the desired (previously
stored) set-up is located.
*SAV<
n
>
Save state
Stores present settings of ARM:COUNt, TRIGger:SOURce,
OUTPut[:STATe], INITiate:CONTinuous, SCAN:MODE, and
SCAN:PORT. <
n
> (0 to 9) is the location in memory where the current
set-up is located.
Note: See the
HP 75000 Series B E1300/E1301 Mainframe User’s Manual
or the ANSI/IEEE Standard 488.2-1987 for
more information about these commands.
70 HP E1368A/69A/70A Command Reference