background image

[ROUTe:]

The

ROUTe

subsystem controls switching and scanning operations for the

Microwave Switch modules in a switchbox.

Subsystem Syntax

[ROUTe:]

CLOSe <

channel_list

>

CLOSe? <

channel_list

>

OPEN <

channel_list

>

OPEN? <

channel_list

>

SCAN <

channel_list

>

:MODE NONE | VOLT
:MODE?

CLOSe

[ROUTe:]CLOSe <

channel_list

>

closes the Microwave Switch channels

specified in the channel_list. The channel_list is in the form (@ccnn)
where cc = card number (00-99) and nn = channel number (00-04).

Parameters

Parameter Name

Parameter Type

Range of Values

Default Value

channel_list

numeric

cc00 - cc04

N/A

Comments

Closing Channels: To close:

a single channel, use

[ROUTe:]CLOSe (@ccnn)

;

multiple channels, use

[ROUTe:]CLOSe (@ccnn,ccnn)

;

sequential channels, use

[ROUTe:]CLOSe (@ccnn:ccnn)

;

a group of sequential channels, use

[ROUTe:]CLOSe (@ccnn:ccnn,ccnn:ccnn)

;

or any combination.

Closure order for multiple channels with a single command is not
guaranteed.

Related Commands:

[ROUTe:]OPEN

,

[ROUTe:]CLOSe?

*RST Condition: All channels are open.

Example

Closing Microwave Switch Channels

This example closes channels 100 and 202 of a two-module switchbox (card
numbers 01-02).

CLOS (@100,202)

100 closes channel 00 of card #1;
202 closes channel 02 of card #2.

58 HP E1368A/69A/70A Command Reference

Chapter 4