Close, Open – HP E1368A User Manual
Page 60
CLOSe?
[ROUTe:]CLOSe? <
channel_list
>
returns the current state of the
channel(s) queried. The channel_list is in the form (@ccnn). The
command returns 1 if the channel is closed or returns 0 if the channel is
open.
Comments
•
Query is Software Readback: The
[ROUTe:]CLOSe?
command
returns the current software state of the channel specified. It does
not account for relay hardware failures. A maximum of 127
channels at a time can be queried for a multi-module switchbox.
Example
Query Channel Closure
This example closes channels 100 and 202 of a two-module switchbox (card
numbers 01 and 02) and queries channel closure. Since the channels are
programmed to be closed 1,1 is returned.
CLOS (@100,202)
100 closes channel 00 of card #1;
202 closes channel 02 of card #2.
CLOS? (@202)
Query channel 02 of card #2.
OPEN
[ROUTe:]OPEN <
channel_list
>
opens the Microwave Switch channels
specified in the channel_list. The channel_list is in the form (@ccnn)
where cc = card number (00-99) and nn = channel number (00-04).
Parameters
Parameter Name
Parameter Type
Range of Values
Default Value
channel_list
numeric
cc00 - cc04
N/A
Comments
•
Opening Channels: To open:
–
a single channel, use
[ROUTe:]OPEN (@ccnn)
;
–
for multiple channels, use
[ROUTe:]OPEN (@ccnn,ccnn)
;
–
sequential channels, use
[ROUTe:]OPEN (@ccnn:ccnn)
;
–
a group of sequential channels, use
[ROUTe:]OPEN (@ccnn:ccnn,ccnn:ccnn)
;
–
or any combination.
Opening order for multiple channels with a single command is not
guaranteed.
•
Related Commands:
[ROUTe:]CLOSe
,
[ROUTe:]OPEN?
•
*RST Condition: All channels are open.
Example
Opening Microwave Switch Channels
This example opens channel 00 of a card number #1 and channel 02 of a
card number #2 in a two-module switchbox.
OPEN (@100,202)
100 opens channel 00 of card #1;
202 opens channel 02 of card #2
Chapter 4
HP E1368A/69A/70A Command Reference 59
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com