beautypg.com

HP RM500SL User Manual

Page 91

background image

0611

RM500SL User’s Guide Version 2.8

Page 91

20.8 Invalid HIT calibration

FastFacts 20.8: Invalid HIT calibration

20.9 No REM ref. mic. detected

FastFacts 20.9: No REM ref. mic. detected

20.10 Invalid REM calibration

FastFacts 20.10: Invalid REM calibration

No REM reference mic. detected.

Check connection on the REM probe microphone.

v Check that a probe microphone assembly is securely plugged into the PROBE

socket to the left of the test chamber.

v Unplug the HIT reference microphone in the test chamber and plug the REM

probe microphone in its place. Try HIT calibration using the reference
microphone of the REM probe microphone in place of the HIT reference
microphone. If this is fails, the REM probe microphone is defective and must
be replaced.

Invalid REM Calibration!

This message may be displayed during the REM calibration procedure. It indicates
that the calibration curve is significantly different from what is expected. It may be
accompanied by a REM reference microphone error.

v Check that a probe microphone assembly is securely plugged into the PROBE

socket to the left of the test chamber.

v Try using a new probe tube. Ensure it is correctly attached and positioned.

v Try the steps in

20.9: No REM ref. mic. detected

.

v Unplug the coupler microphone from the test chamber and plug the REM

probe microphone in its place. Position the open end of the probe tube near
the HIT reference microphone in the test chamber (use putty to hold it) and
close the test chamber lid. Run a Multicurve test at a stimulus level of 70 dB
SPL. The result should have the shape of a normal REM calibration curve. If it
does not, the REM microphone is defective and must be replaced.

Invalid HIT Calibration!

This message may be displayed during the HIT calibration procedure. It indicates
that the calibration curve is significantly different from what is expected. It may be
accompanied by a missing HIT reference microphone error.

v Check that the there is no coupler on the coupler microphone and that the

microphones are correctly positioned for HIT calibration.

v Check that the coupler microphone is securely plugged into the correct socket.

v Follow the suggestions in

20.7: No HIT reference mic. detected

.