7.9
Calibration of HIT reference microphone
FastFacts 7.9: Calibration of HIT reference microphone
Remove any coupler from the coupler microphone before calibration.
1) Position the HIT reference microphone opening concentric with the coupler
microphone opening and about 1 – 2 mm from as shown below.
2) Close and seal the test chamber by tightening the latch at the front of the lid.
3) Press
Instruments Test column.
4) Highlight, then
5) Highlight and
See 7.8: HIT calibration facts
It is recommended that you print a copy of the initial calibration curve and compare
it with your daily or weekly curve. Any change in this curve indicates a drift in
either the coupler or the reference microphone which needs further investigation.
See
7.10: Calibration check for coupler microphone
for an additional check.