14 Occlusion Effect Test
14.1 Occlussion effect measurement
FastFacts 14.1: Occlusion effect measurement
1) Refer to
10: Real-Ear Measurement Setup
to prepare the system for use and
for proper positioning of the client and the probe tube.
2) Insert the earmold or custom hearing instrument into the ear, being careful not
to advance the probe tube further into the ear canal. A lubricant applied to the
earmold or custom instrument shell in the vicinity of the probe tube will make
insertion easier and prevent slit leaks around the tube.
3) Turn the hearing aid OFF.
4) Press
5) Highlight and
Test].
6) While the client vocalizes “ee” in a normal voice, highlight and
test] to stop the test and freeze the screen. A red occlusion bar indicates an
occlusion problem, a yellow bar indicates a possible problem and a green bar
indicates no problem.
7) Increase venting and repeat the test until a green bar is obtained.