10.4 Calibration check for probe module
FastFacts 10.4: Calibration check for probe module
The calibration curve for the probe module is a comparison of the response of the
probe microphone (with tube) to the response of the REM reference microphone.
Any change in the curve indicates a change in one of these microphones (or the
probe tube). See
10.3: Calibration of REM probe microphone
. As part of a regular
calibration schedule, or to investigate changing calibration curves, the REM
reference microphone may be checked against a high quality sound level meter.
1) Press
2) Highlight &
3) Highlight and
drop-down list.
4) Highlight, then
changes to [Stop test]).
5) Hold the probe module about 18” (0.5 m) in
front of the REM speakers.
6) Hold the microphone of a calibrated sound
level meter next to the REM reference
microphone as shown. Set the sound level
meter to C weighting and slow.
7) The sound level meter should read 70 ±2 dB.
8) Repeat with the frequency at 250 Hz.
9) To stop the measurement, highlight and
The probe module should be replaced if it
fails this test.