18.5 Fitting to targets for soft speech

FastFacts 18.5: Fitting to targets for soft speech

See

17.9: REM or S-REM mode

,

17.6: Using Speechmap

and

18.2: Assessment

data entry

for setup and audiometric data entry.

1) Select [Audiometry] to choose the target method and enter audiometric data.

2) Select REAR 1 and set stimulus to [Speech-std(1)] or [Speech-std(2)] and level

to 50. Elongated

+

symbols show the target range for 50 dB speech LTASS.

3) Adjust frequency shaping and gain for soft sounds so that the middle curve (the

LTASS) falls within the target range, especially between 500 - 4000 Hz. Press
to run the complete passage. Repeat as necessary.