HP RM500SL User Manual
Page 61
![background image](/manuals/90282/61/background.png)
0611
RM500SL User’s Guide Version 2.8
Page 61
15.3 RECD coupler response
FastFacts 15.3: RECD coupler response
Prepare the system for HIT tests. See
2.3: Microphone connection
,
7.9:
Calibration of HIT reference microphone
. Connect the optional RECD transducer
to the RECD jack to the left of the test chamber.
1) Press
tests list.
2) If a valid coupler response exists,
go to
15.4:RECD real-ear
response.
3) If no coupler response exists or the
measurement interval has been
exceeded, connect the RECD
transducer to the BTE coupler.
4) Highlight and
5) Press
curve is stable, press
6) Highlight and
the coupler response.
For comparison purposes, an age-related average RECD is shown as a dotted
curve on the screen. If you have accessed RECD from the REM tests list, you may
highlight and