HP RM500SL User Manual
Page 49
0611
RM500SL User’s Guide Version 2.8
Page 49
9.7
Manual control procedure
FastFacts 9.7: Manual control procedure
9.8
Sound level meter using manual control
FastFacts 9.8: Sound level meter using manual control
Manual control displays the reference microphone SPL, the coupler microphone
SPL, their difference in dB (gain) and coupler microphone signal distortion at any
selected 1/12
th
octave frequency and at a selected input level from 40 to 90 dB (in
5 dB steps) and at 0 dB (off).
1) Press
2) Highlight and
window, then highlight and
desired stimulus level from the drop-
down list.
3) Highlight and
window, then highlight and
desired frequency from the drop-down
list. (Distortion measurement data is
available only from 200-4000 Hz.)
4) Highlight and
start the test. [Start test] changes to
[Stop Test]. To stop the measurement,
highlight and
may change Level and Frequency
without stopping the test.
When set for a stimulus level of 0 dB (off), the Manual control mode allows the
RM500SL to be used as a sound level meter. Sound at the HIT reference
microphone and the coupler microphone may be analyzed by 1/12
th
octave, A-
weighted or C-weighted filters. A and C-weighted measurements are limited to
frequencies above 100 Hz and the averaging time and update rate are 384 ms.
1) Press
2) Highlight and
stimulus level from the drop-down list.
3) Highlight and
th
octave, A-weighted, or C-weighted from the drop-down list.
4) For 1/12
th
octave filter, highlight and
highlight and
5) Highlight and
Test]. To stop the measurement, highlight and
change Level and Frequency without stopping the test.
NOTE: A microphone extension cable (VA-130) is available from Audioscan.
Standard audio extension cables should not be used. They will substantially
increase noise levels.