9.3
Multicurve procedure
FastFacts 9.3: Multicurve procedure
9.4
Multicurve results
FastFacts 9.4: Multicurve results
Multicurve generates 1 to 4 output or gain response curves using either pink noise
or swept pure tones at input levels ranging from 40 – 90 dB SPL and 0 dB (off).
1) Press
2) Refer to
7.11: Coupling the hearing instrument
,
7.12: Positioning the hearing
instrument
and
2.5: Battery pill use and care
to prepare the instrument for
testing. Turn the instrument ON.
3) Highlight and
4) Highlight and
[Pink noise] or [Swept] from the drop-down list.
5) Highlight and
and
6) Press
7) To measure another response curve, highlight and
to overwrite a curve, highlight and
To change scale from dB SPL to dB Gain, highlight &
1) Output curves for the noise signal are in 1/12
th
octave bands. They will be
approximately 18 dB lower than pure tone curves for a linear instrument.
2) Gain curves for a linear instrument will be the same for both swept and pink
noise test signals.
3) For compression instruments, both gain and output curves may be expected to
differ for the two test signals.
4) See
7: Hearing Instrument Tests Setup
to change screen appearance.