beautypg.com

KEPCO MST Series User Manual

Page 44

background image

3-12

MST SERIES 061813

4. Issue commands from host computer to read back current.

5. Compare the read back current (step 4) to the measured current (step 3). Verify that mea-

sured current is equal to readback current within the acceptable range of ± 1 LSB (1/4096) of
I

MAX

. If the acceptable range is exceeded, adjust R52 until measured current is within ±0.5

LSB (1/8192) of read back current, repeating steps 1 through 5 as necessary.