8. Verify that DVM connected across pins 5 and 2 of DC Output connector reads 0V.

9. Set switch S1 to CLOSE SENSE position and issue commands from the host computer to

set the MST Power Module to Voltage Mode, program output voltage to –E

MAX

, current limit

to I

MAX

and enable the output.

10.Verify that VOLTAGE MODE, POLARITY REVERSED and OUTPUT ENABLED indicators

on front panel are on and VOLTS meter on front panel indicates E

MAX

.

11. Connect DVM across pins 5 (+) and 2 (–) of the DC Output connector (Figure 2-3) and verify

that DVM reads –E

MAX

.

12.Issue commands from the host computer to read back voltage; verify that readback voltage

is –E

MAX

.

13.Set switch S1 (Figure 2-3) to OPEN SENSE position. Verify that front panel OUTPUT

ENABLED indicator is off, and VOLTS meter reads 0V.

14.Verify that DVM connected across pins 5 and 2 of DC Output connector reads 0V.

15.Set switch S1 to CLOSE SENSE position and issue commands from the host computer to

set the MST Power Module to Current Mode, program output current to I

MAX

and voltage

limit to +E

MAX

.

16.Verify that CURRENT MODE indicator on front panel is on and AMPS meter on front panel

indicates I

MAX

.

17.Issue commands from the host computer to read back current; verify that read back current

is I

MAX

.

SCPI COMMANDS

CIIL COMMANDS

FUNC1:MODE VOLT
VOLT1 –36; CURR1 5; OUTP ON

FNC DCS :CH1 SET VOLT –36 SET CURL 5
CLS :CH1

SCPI COMMANDS

CIIL COMMANDS

MEAS1:VOLT?

FNC DCS VOLT :CH1
INX VOLT
FTH VOLT

SCPI COMMANDS

CIIL COMMANDS

FUNC1:MODE CURR
CURR1 5; VOLT1 36

FNC DCS :CH1 SET CURR 5 SET VLTL 36

SCPI COMMANDS

CIIL COMMANDS

MEAS1:CURR?

FNC DCS CURR :CH1
INX CURR
FTH CURR