Figure 3-1. calibration test set-up, 2 +5 volts reference adjust (r49), 5 volts reference adjust (r49) -9 – KEPCO MST Series User Manual

Page 41: Calibration test set-up -9

background image

MST SERIES 061813

3-9

FIGURE 3-1. CALIBRATION TEST SET-UP

3.8.2

+5 VOLTS REFERENCE ADJUST (R49)

This procedure establishes the +5V reference voltage used by the analog circuits.

1. Disconnect the load and turn on the UUT. Proceed to step 3 if VOLTAGE MODE indicator is

on.