Ac filter, Ac filter -45 – RIGOL DM3068 User Manual
Page 87

Chapter 2 Front Panel Operations
RIGOL
User’s Guide for DM3068
2-45
AC Filter
AC filter applies to ACV, ACI and FREQ/PERIOD measurements. It can optimize the
low-frequency accuracy and minimize the AC settling time. DM3068 provides three
types of AC filters (slow, medium and fast). The AC filter to be used is determined
by the input signal frequency.
Input Frequency
AC Filter
3 Hz to 300 kHz
Slow
20 Hz to 300 kHz
Mid
200 Hz to 300 kHZ
Fast
In ACV, ACI or FREQ/PERIOD measurement, press
Filter to enter the
following interface.
Press Slow, Mid or Fast to select 3 Hz, 20 Hz or 200 Hz as the lowest frequency of
the AC filter. The default is 20 Hz: MEDIUM.