Measurement configuration, Measurement configuration -40 – RIGOL DM3068 User Manual
Page 82

RIGOL
Chapter 2 Front Panel Operations
User’s Guide for DM3068
2-40
Measurement Configuration
Most measurement parameters are user-defined. Changing a measurement
parameter will change the measurement precision and speed as well as the input
impedance. An appropriate measurement parameter based on the actual
application will ensure faster measurement or higher measurement precision.
The default measurement configurations of the multimeter can ensure the
accuracy of the measurement results in most cases. Users can directly use these
defaults for any measurement or modify the parameters of the measurement
function as required.
The parameters for different measurement function differ, see table below.
Table 2-5 Measurement parameter
Functions
Parameters
DCV
integration time, DC impedance, Auto zero (AZ)
ACV
AC filter
DCI
integration time, Auto zero (AZ)
ACI
AC filter
OHM (2WR, 4WR)
Integration time, Auto zero (AZ), Offset compensation
(OC)
CAP
N/A
CONT
Short-circuit resistance
DIODE
Current Source
FREQ/PREIOD
Gate time, AC filter
SENSOR
N/A
To set the measurement parameter, select a measurement function and then press
. For more details, please refer to the following content.