Model PD6300 Pulse Input Rate/Totalizer

Instruction Manual

80

Meter Copy or Cloning Instructions

!

Caution!

Do not connect the two meters to
the same signal source while
cloning. Internal calibration may
be affected.

1. Connect two meters using a PDA1200 meter copy cable.

Warning!

Use PDA1200 meter copy cable only!
Using standard CAT5 or other cable will cause
damage to both meters.

2. Do not connect the two meters to the same signal source.
3. Power up both meters. Leave Clone meter in Run Mode.
4. Enter

the

Advanced Features

menu of the Master meter see Ad-

vanced Features Menu

page 70.

5. Scroll

to

the

Copy

function using the Up arrow button then press

Enter.

6. The meter displays the message Send. Press Enter, the display

flashes while sending data. The message done is displayed when
copying is completed.

Send

Copy

done

Copy

Display flashes

while sending

data

Copy

7. The Clone meter displays the message CPY rC while being pro-

grammed then the message done when copying is completed. The
meter initializes and returns to Run Mode using the same settings
as the Master.

8. If meter to be cloned does not respond to the data being sent, refer

to Copy Function Requirements above.