Installation, Unpacking, Panel mounting instructions – Precision Digital PD6300 User Manual
Page 17: Figure 1: 1/8 din panel cutout and mounting
Model PD6300 Pulse Input Rate/Totalizer
Instruction Manual
17
INSTALLATION
There is no need to remove the meter from its case to complete
the installation, wiring, and setup of the meter for most applica-
tions. Instructions are provided for setting up a 12/24 VDC po-
wered meter to operate from 12 VDC and for changing the
transmitter power supply to output 5 or 10 VDC instead of 24
VDC, see page 19.
Unpacking
Remove the meter from box. Inspect the packaging and con-
tents for damage. Report damages, if any, to the carrier.
If any part is missing or the meter malfunctions, please contact
your supplier or the factory for assistance.
Panel Mounting Instructions
Prepare a standard 1/8 DIN panel cutout – 3.622" x 1.772" (92 mm x
45 mm). Refer to Figure 1 for more details.
Clearance: allow at least 6.0" (152 mm) behind the panel for wiring.
Panel thickness: 0.04" - 0.25" (1.0 mm - 6.4 mm).
Recommended minimum panel thickness to maintain Type 4X
rating: 0.06" (1.5 mm) steel panel, 0.16" (4.1 mm) plastic panel.
Remove the two mounting brackets provided with the meter (back-off
the two screws so that there is ¼" (6.4 mm) or less through the brack-
et. Slide the bracket toward the front of the case and remove).
Insert meter into the panel cutout.
Install mounting brackets and tighten the screws against the panel.
To achieve a proper seal, tighten the mounting bracket screws
evenly until meter is snug to the panel along its short side. DO NOT
OVER TIGHTEN, as the rear of the panel may be damaged.
3.622" (92mm)
1.772"
(45mm)
Panel Cutout
to DIN 43700
Square Corners to 0.060"
(1.5mm) Max Radius
A
B
Tolerances:
A: +0.032 (+0.8mm)
-0.000 (-0.0mm)
B: +0.024 (+0.6mm)
-0.000 (-0.0mm)
Panel
Gasket
Mounting
Bracket
Mounting
Screw
Removable
Connectors
Figure 1: 1/8 DIN Panel Cutout and Mounting