Model PD6300 Pulse Input Rate/Totalizer

Instruction Manual

43

K-Factor Calibration (Factor)

The meter may be calibrated using the K-Factor function. Most flowmeter
manufacturers provide this information with the device. Enter the
K-Factor

(Factor) menu and select the decimal point with highest reso-

lution possible and program the k-factor value (i.e. pulses/gal). The me-
ter will automatically calculate the flow rate using the k-factor and the
time base selected.

ddd.ddd

factor

Factor

prog

d.ddddd

factor

1.8925

6

factor

Multi-Point Calibration & Scaling

The meter is set up at the factory for 2-point linear calibration. The num-
ber of points for multi-point calibration/scaling is set up in the Advanced
Features

menu. Up to 32 linearization points may be selected. See page

76 for details.