Specifications – LW Scientific OmniVID User Manual

Page 26

background image

25

calibration. The scale on the bottom left corner will be changed accordingly.

The three circles rights next to the three buttons of the Measure Function dialogue

box are used to select the color of the measurement tools during calibration. You

can choose a different color according to the image background by clicking the

circles.

The calibration with a certain objective can only be used for the measurement with

the same objective. When a different objective is used, a new calibration must be

done for measurement.

Line:

When the calibration is done, you can measure the distance between two points on

the preview screen by clicking Line button. Use mouse click to select two points on

the screen, and the real distance will be

shown in the dialogue automatically.

Circle:

You can also measure the circumference and radius of a circle on the screen if the

measurement is properly calibrated. Use mouse click to select three points on the

screen, and the dialogue will show the circumference and radius of the circle

defined by these three points.

The lines and circles drawn on the screen for measurement will not be recorded in

the image.

Specifications

Camera

t

n

u

o

M

-

C

t

n

u

o

M
Sensor resolution :

3.1 Mega Pixel

Maximum resolution :

8 Mega Pixel

LCD screen size

2 inch LTPS

Memory card

SD/MMC card

USB interface

Yes

TV output

NTSC/PAL

Controller

Remote and USB shutter

Power supply

DC 12V 1.5A

Picture format

JPEG

Video format

MPEG4

Printer support

PictBridge and DPOF

PC Software

MicroCap

V2.0