LW Scientific i4 12 Volt Infinity User Manual
I4 12vdc infinity, Instruction manual, Upgrades
Camera Attachments
A
K
B
P. 770.270.1394
F. 770.270.2389
865 Marathon Parkway Lawrenceville GA 30046
P. 770.270.1394
F. 770.270.2389
865 Marathon Parkway Lawrenceville GA 30046
Instruction Manual
E
Brightness Control
F
Focus Friction Control
G
Coarse Focus
H
Fine Focus
I
On/Off Switch
(located on rear)
J
X/Y Axis Stage Controls
K
Base Condenser
L
Filter Holder
M
Substage Iris Diaphragm
N
Substage Abbe Condenser
O Stage
P
Slide Holder
Q Objectives
R Nosepiece
S Eyepieces
A
Diopter Adjustment
B
Binocular Head
C
Head Retention Screw
D
Stage Stop Lever
(located on left side)
Specifications
Maintenance
Ask your
authorized LWS
dealer about
additional
accessories
Upgrades:
Recommended
i4 12vDC Infinity
i4 12vDC Infinity
Introduction
The LW Scientific I-4 microscope features exceptional optical quality with an infinity optical system, like found on the most
expensive laboratory microscopes, but at a fraction of the price. The ergonomic narrow design allows users to rest their
arms flat on the table and easily operate all controls. The binocular eyetubes can also be rotated upward to achieve an
additional 2 inches of height for taller users. LED lighting produces “daylight” color, with a cool temperature and long life.
Comfort, durability, dependability, and superior imaging make the I-4 a great value for labs, physicians, vets, and
universities. The I-4 microscope is intended for use as a biological microscope in a professional environment in accordance
with the guidelines set forth in this operations manual.
2
4
5
3
1
Always cover your microscope with the dust cover when not in use. When cleaning the lenses, use lens paper or a
cotton swab dipped in lens cleaning solution.
Excess oil should be cleaned off your 100x objective and stage at once. An alcohol pad is best for removing oil from
the stage and on the other metal parts, but is not recommended for use on the lenses. Use lens cleaning solution and
lens paper to clean off your objectives.
Dust in the nosepiece or ocular tubes should be blown out using filtered air. Canned air dusters work well for this job.
Whenever you remove an objective, we recommend that you place it back into the original plastic shipping vial until
ready to be placed back on the microscope. SCREW THE OBJECTIVE SECURELY INTO THE CAP OF THE HOLDER - DO NOT
DROP OBJECTIVE LOOSELY INTO CONTAINER.
To keep your microscope in top condition for years, LW Scientific recommends that you have the microscope
professionally serviced once a year.
Warning: The 40x objective is not sealed for oil immersion. Damage to the 40x objective due to oil
immersion is not covered under warranty.
Head
Binocular (Seidentopf)
Inclined 30°, rotates 360°
Diopter adjustment
10X/18 wide-field eyepieces
10X/20 High-Point super-wide eyepieces available
Trinocular available
Nosepiece
Reverse quadruple hole multiple ball bearing
Illumination
Moveable Abbe Condenser (NA 1.25)
Iris Diaphragm
Blue and green filters
LED light
Construction
Rugged alloy
Stage
Mechanical stage (142 mm x 133 mm)
– coaxial drive controls
Range of traverse is 73 mm x 43 mm
Acid and reagent resistant finish
Objectives
Infinity Semi-Plan, Infinity Plan Achromat Objectives
4X N.A. 0.10
10X N.A. 0.25
40XR N.A. 0.65
100XR N.A. 1.25 (oil immersion)
50X oil Plan Infinity available
Adjustment Controls
Eyepiece: Interpupillary distance adjustment 48-75 mm
Stage Controls: Knobs allow movement of slide on X-Y axis
Etched vernier scales
Coarse Adjustment: Range of 30 mm
Fine Adjustment: Graduation of 2μm
Variable Light Adjustment
Dimensions and Weight
Weight: 16 lbs / 7.3 kg
Height: 14.1" / 267 mm
Length: 10.7" / 267.5 mm
Width: 6.1" / 152.5 mm
Shipping weight: 17 lbs/ 8kg
Size
N.A.
Mag. Field of View
4X
0.10
40X
4.5mm
10X
0.25
100X
1.8mm
20X
0.40
200X
0.9mm
40XR
0.65
400X
0.45mm
50XR
0.95
0.36mm
500X
60XR
0.85
0.3mm
600X
100XR 1.25
0.18mm
1000X
Objectives: The following numbers are
based on use with 10x/18 eyepieces.
I
D
C
G
H
F
J
E
L
M
N
O
S
P
Q
R
Lumin
Epi-Module
Objectives
Phase and Dark Field
Operating Environment
Indoor Use Only
Ambient temperature: 5˚ to 40˚C (41˚ to 104˚F)
MKT-7.5.3-L-103 | Rev 3